| Measuring Targets |
Strain gage and strain-gage transducers |
| Channels |
4 |
| Input Form |
Balanced differential input |
| Synchronous Measurement |
Up to 4 units by 16 channels |
| Compatible Gage Resistance |
Quarter bridge 2-wire system, 3-wire system: 120 Ω
Half bridge system, full bridge system: 120 to 1000 Ω |
| Input Connector |
D-sub 37-pin connector |
| Bridge Excitation |
AC 2 Vrms |
| Gage Factor |
2.00 fixed |
| Balance Adjustment |
Resistance: Within ±2% (±10 k ×10-6 strain)
Capacitance: Within 5000 pF |
| Balance Adjustment Methods |
Resistance: Auto balance
Capacitance: CST method (Capacitance self-tracking) |
| Nonlinearity |
Within ±0.1% FS |
| Range |
200, 500, 1 k, 2 k, 5 k, 10 k, and 20 k ×10-6 strain – 7 steps
Accuracy: within ±0.5% FS |
| Frequency Response |
DC to 200 Hz, deviation: Within ±10% |
| Sampling Frequencies |
Max. 10 kHz (Synchronous 4-unit sampling for 16 channels at 10 kHz) |
| LPF |
Transfer characteristic: 2nd order Butterworth
Cutoff frequencies: 10, 30, 100 Hz, and FLAT (4 steps)
Amplitude ratio at cutoff point: -3 ±1 dB
Attenuation: (-12 ±1) dB/oct. |
| AD Converter |
24 bits |
| Holding Setting Condition |
The range and balance adjustment value etc. are written to nonvolatile memory. |
| TEDS |
Reads information from TEDS-installed sensors.
(Input adapters: UI-10A and UI-11A only)
Channel name writing
(If the manufacturer’s ID is Kyowa) |
| Interfaces |
USB2.0 (Conforms to High-speed USB standards.USB3.0 supported) |
| Stability |
Temperature
Zero point: Within ±0.2 ×10-6 strain per ºC
Sensitivity: Within ±0.05%/ºC
Time
Zero point: Within ±0.5 ×10-6 strain per 8h
Sensitivity: Within ±0.15%/8h |
| Withstand Voltage |
250 VAC for 1 minute between input and case
250 VAC for 1 minute between input channels |
| Operating Temperature |
0 to 40ºC |
| Operating Humidity |
20 to 85% (Non-condensing) |
| Vibration Resistance |
±29.42 m/s2 (3 G)
5 to 200 Hz (12 cycles for each axis, 10 minutes/cycle)
*unstacked |
| Power Supply |
11 to 16 VDC
Connector type: RM12BRD-4PH (Hirose) |
| Current Consumption |
0.7 A or less (12 VDC)
0.6 A or less (15 VDC) |
| Dimensions |
210 W x 35 H x 157.5 D mm (Excluding protrusions) |
| Weight |
Approx. 650 g |
| Compliance |
Directive 2011/65/EU, (EU)2015/863 (10 restricted substances) (RoHS)
Directive 2014/30/EU (EMC) |
| Degree of Protection |
IP30 (JIS C 0920 / IEC 60529)
*Except when the one-touch type input adapter UI-55A is connected and when the adapter is not connected.
*The specifications apply to the state in which the temperature has stabilized after a preheating time of 1 hour. |
| Standard Accessories |
USB cable N-38 (1 m)
Ground wire P-72 (5 m)
Dynamic data acquisition software DCS-100A (DVD) |
| Optional Accessories |
AC adapter UNI360-1540-AC
USB cable N-39 (2 m)
DC power cable P-76 (2 m)
Connection cable N-97 (10 cm)
Stacking kit ST-1B
Input adapter for strain gage transducer UI-10A (TEDS compatible)
Input adapter for strain gage UI-11A (TEDS compatible)
Input adapter for strain gage with operating lever UI-15A
One-touch type input adapter for strain gage UI-16B
One-touch type input adapter UI-55A
Data analysis software DAS-200A |