| Measuring Mode |
Strain measurement mode |
Voltage measurement mode |
| Measuring Targets |
Strain gages and strain-gage transducers |
Voltage |
| Channels |
4 |
|
| Input Form |
Balanced differential input |
Unbalanced |
| Synchronous Measurement |
Up to 4 units by 16 channels |
|
| Compatible Gage Resistance |
Quarter bridge 2-wire system, 3-wire system: 120 Ω Half bridge system, full bridge system: 120 to 1000 Ω |
– |
| Input Connector |
D-sub 37-pin connector |
|
| Bridge Excitation |
AC 2 Vrms |
– |
| Gage Factor |
2.00 fixed |
– |
| Balance Adjustment |
Resistance: Within ±2% (±10 k ×10-6 strain) Capacitance: Within 5000 pF |
– |
| Balance Adjustment Methods |
Resistance: Auto balance Capacitance: CST method (Capacitance self-tracking) |
– |
| Nonlinearity |
Within ±0.1% FS |
– |
| Range |
200, 500, 1 k, 2 k, 5 k, 10 k, and 20 k ×10-6 strain, -7 steps Accuracy: Within ±0.5% FS |
1, 2, 5, 10, 20, and 50 V -6 steps Accuracy: Within ±0.2 %FS |
| Frequency Response |
DC to 200 Hz, Deviation: Within ±10% |
DC to 1 kHz, Deviation: Within -3 to 1 dB |
| Sampling Frequencies |
Max. 10 kHz (Synchronous 4-unit sampling for 16 channels at 10 kHz) |
|
| LPF |
Transfer characteristic: 2nd order Butterworth Cutoff frequencies: 10, 30, 100 Hz, and FLAT (4 steps) Amplitude ratio at cutoff point: -3 ±1 dB Attenuation: (-12 ±1) dB/oct. |
2nd order Butterworth Cutoff frequency: 10, 30, 100, 300 Hz, and FLAT (5 steps) Amplitude ratio at cutoff point: -3 ±1 dB Attenuation characteristics (-12 ±1) dB/oct. |
| HPF |
– |
0.2 Hz or OFF (2 steps) |
| AD Converter |
24 bits |
|
| Setting Value Storage |
– |
The range and balance adjustment value etc. are written to nonvolatile memory. |
| TEDS |
Strain measurement mode: Reads information from TEDS-installed sensors. (Input adapters: UI-10A and UI-11A only) Channel name writing (If the manufacturer’s ID is Kyowa) |
– |
| Interfaces |
USB2.0 (Conforms to High-speed USB standards. USB3.0 supported) |
|
| Stability |
Temperature Zero point: Within ±0.2 ×10-6 strain per ºC Sensitivity: Within ±0.05%/ºC Time Zero point: Within ±0.5 ×10-6 strain per 8h Sensitivity: Within ±0.15%/8h |
Temperature Zero point: Within ±0.008 %FS/ºC Sensitivity: Within ±0.02 %/ºC Time Zero point: Within ±0.03 %FS/8h Sensitivity: Within ±0.1 %/8h |
| Withstand Voltage |
Strain measurement mode: 250 VAC for 1 minute between input and case 250 VAC for 1 minute between input channels |
– |
| Operating Temperature |
0 to 40ºC |
|
| Operating Humidity |
20 to 85% (Non-condensing) |
|
| Vibration Resistance |
±29.42 m/s2 (3 G) 5 to 200 Hz (12 cycles for each axis, 10 minutes/cycle) *unstacked |
|
| Power Supply |
11 to 16 VDC Connector type: RM12BRD-4PH (Hirose) |
|
| Current Consumption |
0.9 A or less (12 VDC) 0.7 A or less (15 VDC) |
|
| Dimensions |
210 W x 35 H x 157.5 D mm (Excluding protrusions) |
|
| Weight |
Approx. 670 g |
|
| Compliance |
Directive 2011/65/EU, (EU)2015/863 (10 restricted substances) (RoHS) Directive 2014/30/EU (EMC) |
|
| Degree of Protection |
IP30 (JIS C 0920 / IEC 60529) *Except when the one-touch type input adapter UI-55A is connected and when the adapter is not connected. |
|
|
* The specifications apply to the state in which the temperature has stabilized after a preheating time of 1 hour. |
|
| Standard Accessories |
USB cable N-38 (1 m) Ground wire P-72 (5 m) Dynamic data acquisition software DCS-100A (DVD) |
|
| Optional Accessories |
AC adapter UNI360-1540-AC USB cable N-39(2 m) DC power cable P-76 (2 m) Connection cable N-97 (10 cm) Stacking kit ST-1B NDIS-BNC Conversion adapter FV-1A Input adapter for strain gage transducer UI-10A (TEDS compatible) Input adapter for strain gage UI-11A (TEDS compatible) Input adapter for strain gage with operating lever UI-15A One-touch type input adapter for strain gage UI-16B One-touch type input adapter UI-55A Voltage input adapter UI-30A Data analysis software DAS-200A |
|