UCAM-60C M14 – Data Logger
Summary
The UCAM-60C M14 is an all-in-one measuring instrument
The data logger UCAM-60C M14 is an all-in-one measuring instrument developed in full pursuit of easier field measurement. Easy to operate keys, a bright readable display providing understandable presentation and a printer for immediate confirmation of measurement results. All these and more are incorporated in this compact unit to satisfy every need in field measurement.
The UCAM-65C M14 is an online data logger fully controlled from the PC.
The UCAM-60C M14 is an all-in-one measuring instrument
Model AC power supply models: UCAM-60C-AC M14
DC power supply models: UCAM-60C-DC M14
Channels Max 30 with dedicated scanners
Max 1000 with external scanners connected
Max 1000 with dedicated scanners and external scanners connected
Measuring Targets and Connectable Scanners
Measuring Targets Strain gages, strain-gage transducers, civil engineering transducers with a thermal sensor, DC voltage-output or DC current-output instruments,
potentiometer sensors, thermal sensors (Thermocouples and platinum resistance thermometer bulbs)
Connectable Scanners USS-61B, USS-62B, USS-63B (Dedicated Scanners, mounted on top of the UCAM-60C)
The main unit can connect scanners USB-70B series via an optional scanner interface USI-67A
Input Terminals Capable of connecting lead wires through either soldering or screwing. (Screw: M3 with metal plate)
NDIS connectors (USS-62B)
Switching Terminals Semiconductor relays
Scanning Speed 50 ms/channel (Standard mode)
0.28 s/channel (High-resolution mode)
Note 1: Individually switchable for desired channels.
20 ms/channel (High-speed mode)
Note 2: Only collective switching for all channels of dedicated scanners.
Note 3: High-resolution mode and high-speed mode are selectable for dedicated scanners only.
Note 4: High-resolution mode is available only with full bridge system.
Note 5: High-speed mode is available full bridge system, voltage, current, and potentiometer sensor.
Operating Modes Real-time, monitor, and automatic
Measurement Functions
Initial Initial values are measured and stored in internal memory.
Original Raw values are measured without subtraction of initial values.
Measure Initial values are subtracted from original values.
Easy Measure Auto zero balancing function is activated.
Notes The selected function is applied to all channels.
Coefficient Calculation Function Multiplication by calibration coefficients, calibration by TEDS, conversion of measured values to physical quantities, scaling and correction.
Units 59 units
Automatic Measurement Function
Trigger Measurement A relative value (Certain changing quantity) or an absolute value trigger measurement.
Interval Measurement Measurement is automatically performed at preset time intervals.
Trigger Interval Measurement Combination of trigger measurement and interval measurement.
Storage Internal memory, approx. 1.8 GB
Strain Measurement (Standard Mode)
Constant Voltage Excitation Approx. 2 or 5 VDC
Constant Current Excitation Approx. 5.7 mA (Bridge resistance 350 ?)
Approx. 16.7 mA (Bridge resistance 120 ?)
Scanning Speed 50 ms/channel
Gage Factor 2.00 fixed
(Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
0 to ?50k ?10-6 strain
?50k to ?500k ?10-6 strain
Strain Measurement (High-resolution Mode)
Constant Voltage Excitation Approx. 5 VDC
Constant Current Excitation Approx. 16.7 mA (Bridge resistance 350 ?)
Scanning Speed 0.28 s/channel
Gage Factor 2.00 fixed
(Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
0 to ?20k ?10-6 strain
?20k to ?200k ?10-6 strain
Notes 1. Available only with full bridge system
2. Bridge resistance should be 120 to 1000 ? for bridge excitation with constant voltage.
3. Bridge resistance should be 350 ? for bridge excitation with constant current.
Measuring range is 0 to 15k ?10-6 strain, 0 to 150k ?10-6 strain
4. Available only with dedicated scanners.
Strain Measurement (High-speed Mode)
Constant Voltage Excitation Approx. 2 VDC
Constant Current Excitation Approx. 5.7mA (Bridge resistance 350 ?)
Approx. 16.7mA (Bridge resistance 120 ?)
Scanning Speed 20 ms/channel
Gage Factor 2.00 fixed
(Coefficient calculation function enables correction with 2.00/Ks)
Initial Value Memory Range Same as measuring range
0 to ?50k ?10-6 strain, ?50k to ?500k ?10-6 strain
Notes 1. Available only with full bridges system (120 to 1000 ?)
2. Available only with dedicated scanners.
Voltage Measurement (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Range Same as measuring range
V/500mV Range Mode: 0 to ?50.000 mV, ?50.00 to ?500.00 mV
V/50V Range Mode: 0 to ?5.0000 V, ?5.000 to ?50.000 V
Voltage Measurement (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
V/500mV Range Mode: 0 to ?50.000 mV, ?50.00 to ?500.00 mV
V/50V Range Mode: 0 to ?5.0000 V, ?5.000 to ?50.000 V
Notes 1. Resolution and accuracy be automatically change by autorange function.
2. Available only with dedicated scanners.
Current Measurement (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Range Same as measuring range
0 to ?50.00 mA
Notes External shunt resistors (High-accuracy 250 ? are required.
Current Measurement (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
0 to ?50.00 mA
Notes 1. Available only with dedicated scanners.
2. External shunt resistors (High-accuracy 250 ? are required.
Temperature Measurement with Thermocouple (Standard Mode)
Scanning Speed 50 ms/channel
Notes Thermocouple resistance should be 1000 ? or less.
Temperature Measurement with Civil Engineering Transducers with a Thermal Sensor (Standard Mode)
Scanning Speed 50 ms/channel
Notes 1. Target physical quantity and temperature are measured in a single channel.
2. Strain measuring range is the same as in strain measurement in standard mode.
Temperature Measurement with Platinum Resistance Thermometer Bulb (Standard Mode)
Scanning Speed 50 ms/channel
Measurement with Potentiometer Sensor (Standard Mode)
Scanning Speed 50 ms/channel
Initial Value Memory Same as measuring range
0% to ?50.00%
Sensor Power Supply Approx. 2 VDC
Potentiometer Resistance 1 to 10 k ?
Measurement with Potentiometer Sensor (High-speed Mode)
Scanning Speed 20 ms/channel
Initial Value Memory Range Same as measuring range
0% to ?50.00%
Sensor Power Supply Approx. 2 VDC
Potentiometer Resistance 1 to 10 k?
Notes Available only with dedicated scanners.
Internal Timer Real time clock is Built-in (Battery backup)
Display Fluorescent display tube,
128?64 dots
Printer Printing: Thermal
Paper Width: 58 mm (24 characters/line)
Printing Speed: 60 mm/s
Interfaces RS-232C
LAN (10BASE-T, 100BASE-TX?
USB2.0 (Collect measurement data by USB Flash Drive)
Notes 1: Measurement data cannot be saved directly to USB memory.
Notes 2: USB Flash Drives Capacity is 32 GB or less
File Format: FAT32
Notes 3: Recommend USB Flash Drives GH-UFI-XSC2G (The manufacturer is Greenhouse)
File Conversion Binary measurement data can be converted to CSV.
Self-diagnosis Functions Checks display, printer, bridge excitation, lead-wire-off, I/O resistance, insulation resistance, mode, etc.
TEDS ?Interface
IEEE1451.4 Mixed mode Transducer Interface Class2
?Applicable sensor
Should have information written in accordance with IEEE template No.33
Cable length should be 30m or less
(With dedicated scanners USS-61B,USS-62B, USS-63B)
Operating Temperature 0 to 50 ?C
Operating Humidity 20% to 85%RH (Non-condensing)
Storage Temperature -20 to 60 ?C
Power Supply AC 100 to 240 V (AC-operated models)
DC 10 to 16 V (DC-operated models)
Note: DC operated models have power control function
Current Consumption 0.5 A or less: 100 VAC (With 3 dedicated scanners mounted)
4.0 A or less: 12 VDC (With 3 dedicated scanners mounted)
Dimensions 360 W ?88 H ?400 D mm (Excluding protrusions)
Weight Approx. 6.3 kg (Excluding scanners)
Approx. 9.6 kg (With 3 dedicated scanners USS-62B mounted)