WGC-140A – 4-channel Instrumentation Amplifier

Kyowa

The WGC-140A is a high-speed processing at 2 ms, a compact 4 -channel instrumentation conditioner, which can calculate the physical quantities of load cells, etc. It is suitable for measurement and control of the load applied to each column of press and the total load or for tension control by measurement of load balance and total load in various fields.

Kyowa

WGC-140A – 4-channel Instrumentation Amplifier

Summary

  • Up to 4 units of transducers can be connected independently to the respective amplifier circuits.
  • Switchable indicated values of channel to channel or the total
  • Analog voltage output of each channel and the total
  • High/low limit comparators for the total values
  • High-speed output suitable for control, I/O delay approx. 10 ms
  • Sensor check and self-test functions
  • Display -19999 to 19999 with decimal point anywhere

Technical Specifications

Channels Max.4
Applicable Transducers Strain-gage transducers
Applicable Bridge Resistance 350 to 1000 ?
Measuring Range ?3 mV/V (Including tare)
Input LPF 150 Hz
Bridge Excitation 5 VDC
Digital Zero Set at any point in the measuring range
Calibration By inputting the value (By every channel)
By applying an actual load (By every channel)
Indicator ?19999 (Decimal point to be put anywhere, the same position is applied to all channels.)
Character height: 14 mm, red LED
Indicated value: Measured value of a channel selected from 1 to 4
Total of measurements of channels 1 to 4
Calculation Speed 2 ms (500 times/s)
Nonlinearity Within ?0.05% FS
Stability Zero point: Within?0.5 ?VRTI/?C
Sensitivity: Within?0.01%/?C
High/Low Limit Comparator Setting points: 2 (High/low limits of total value)
Setting range: ?19999
Max. hysteresis width: 0 to 19999 settable
Output modes: Open collector
Load capacity: 30 VDC, 20 mA (Resistive load)
Response speed: 10 ms or less
Smoothing Functions Minimum scale: Selectable from 1, 2, 5, 10, 20, 50, 100, 200, 500 or 1000 counts
Moving averaging functions: Selectable from 2, 4, 8, 16, 32, 64, 128 or 256 times
Adding Functions Setting range: ?19999
Original Value Measurement ?3 mV/V or more
Control Input Input signals: 6 (4 calculation channel select commands, 1 each ZERO and CHECK commands)
Input modes: Non-voltage contact or open collector (It should enable application of 12 VDC and current flow of 5 mA.)
Control Output Output signals: 7 (1 HEALTHY signal, 4 ABNORMAL signals and high/low limit signals)
Output modes: Open collector
Load capacity: 30 VDC, 20 mA (Resistive load)
Analog Output Output signals: 5 (Signals of 4 channels and the total)
Output voltage: ?10 V
Withstand voltage: 250 VAC for one minute
Resolution: 13 bits
Nonlinearity: Within ?0.1% FS
Conversion rate: 500 times/s
Setting values: Indicated value with 0 V output (?19999)
Indicated value with 10 V output (?19999)
Check Functions Self-test: Tests the program checksum and memory.
Transducer test: Checks each channel for the bridge current, over-input and disconnection of transducer cable.
I/O Terminal Board Transducer input: NDIS connector plugs
Power connector, etc.: M3 screw terminal board (Applicable crimp-style terminal V1.25-3 or the equivalent)
Data output terminal: Connector 57-40360 (DDK) or the equivalent
Operating Temperature -10 to 55?C
Operating Humidity 20 to 85%RH (Non-condensing)
Power Supply 100 to 240 VAC, approx. 30 VA or less
Dimensions 96 W ? 96 H ?139 D mm (Excluding protrusions)
Weight Approx. 1.2 kg
Panel-cut Dimensions 92?92 mm
Standard Accessories Instruction manual, unit seal
Optional Accessories AC power cables
P-23 (For 100 VAC)
P-28 (For 200 VAC)

 

Others

N/A